Search keywords:
product name, product type, model number,
test method, manufacturer, technique, application
Salt Aerosol Automated Filter Tester
BACKStandards:
Applications:
Product Information:
The Salt Aerosol Automated Filter Tester is designed to test high-efficiency filter based products using a sodium chloride salt aerosol. The aerosol particle size and distribution are designed to meet all requirements as defined in the relevant sections of NIOSH 42 CFR, Part 84 and is also suitable for testing to JMOL and in-house QC to EN143/related standards. The 100S can be utilized to meet various flat sheet media, filter, and respirator/mask testing requirements while offering unmatched customer features and benefits. A high level of performance coupled with low service and maintenance requirements makes the 100S ideally suited to meet the rigorous quality control and validation requirements of both development and production quality control environments.
Product Classification:
can be supplied in a manual, automatic, or hybrid mode of operation.
The manual version is ideal for lab use or production lot testing and requires an operator to engage a test using ergonomic photo-electric actuators. Included standard is a manual test fixture and standard 100cm2 flat sheet media fixture.
The automatic version integrates seamlessly with a customer designed, PLC driven, automated production line and is ideal for 100% in-line non-destructive product testing. The unit becomes a slave device to the automated production line and is controlled via communication port. In fact, the 100S is the only salt aerosol filter tester capable of seamless automated production integration.
The Hybrid version provides both the ability to operate in an automated production environment as well as in a manual mode.
Technical Parameters:
Items | Parameters |
Size | 71cm W x 69cm D x 89cm H (28” W x 27” D x 34” H) |
Weight | 75Kg (165bs) |
Aerosol Size & Distribution | 0.075μm Count Median Diameter / 0.26μm Mass Mean Diameter / Geometric Standard Deviation < 1.86 |
Concentration | 35 to 50 milligrams per cubic meter (mg/m3) |
Efficiency | up to 99.9995% |
Pressure | 0 - 200 millimeters of water column (0 - 7.88 inches of w.c.) / accuracy ± 2.5% of full scale |
Flow Rate | 10 to 100 liters per minute (slpm or alpm) / accuracy ± (0.4% of reading + 0.2% full scale) |
Electrical | 110VAC, 60 Hz, 5.5A or 220VAC, 50Hz, 2.5A |
Compressed Air | 566 SLPM @ 5.6 kg/cm2 (20 SCFM @ 80 PSIG) |