Search keywords:

product name, product type, model number,
test method, manufacturer, technique, application

Filter Media Tester G320

BACK
Share

Standards:

Applications:

Testing of complete filters better than ISO 12500
Testing of filter media
Determination of the drainage amount during burdening
 

Product Information:

Compressed air filter test rig
The newly developed compressed air filter test rig delivers fully-automated measurements of fraction separation efficiency for compressed air filters under overpressure up to 10 bar, offering better results in the process than ISO 12500. On request we can also supply an additional filter holder for flat filter media. The fraction separation efficiency is measured with the new aerosol spectrometer Promo® 3000 P, which is equipped with pressure-resistant welas®aerosol sensors for raw gas and clean gas measurements. The system offers the following advantages (among others):
Clear determination of the particle size and concentration directly under overpressure up to 10 bar
Automatic adjustment of the sampling volume flow in the event of pressure changes
No diffuser, no dilution required
The can be equipped with three different aerosol generators:
For measurement of the fraction separation efficiency with oil: PLG 3000
For burden tests, short burdening times with oil: AGF 3000
For tests with solid particles up to 3 bar: RBG 1000 D
The largely automated setup of the test sequence together with the clearly defined individual components and the individually adjustable sequence programs of the filter test software FTControl deliver the high reliability of our measurement results.
All air volume flows and the overpressure in the system are automatically monitored and adjusted. The results are presented in detailed form in tables and diagrams and analyzed. With more than 120 complete filter test systems manufactured and delivered in over 16 different versions, Palas®is the world-leading 

Benefits:
Particle measurements at a glance
Internationally comparable measurement results
High reproducibility of the testing method
Easy use of different test aerosols, e.g. DEHS, SAE Fine and Coarse
Flexible filter test software FTControl
Flexibly programmable sequence programs for pressure loss measurements, measurements of fraction separation efficiency and burden measurements
Fully-automatic and reproducible test sequence including pressure and volume flow control
Easy to operate, even untrained personnel can be instructed quickly in the use of the equipment
Short set-up times
Cleaning and calibration can be performed autonomously by the customer
Easy use of the measurement technology components – even in other applications
Mobile setup, easy to move on castors
Reliable operation
Validation of the clear function of individual components and the overall system during pre-delivery acceptance testing and upon delivery
Low-maintenance
The unit will reduce your operating costs
 
Technical Parameters:

Measurement range (size) 0.2 – 40 µm, 0.3 – 40 µm (at 7 barg)
Volume flow 1 – 60 Nm³/h or 10 – 200 Nm³/h (pressurized operation)
Power supply 115/230 V, 50/60 Hz
Dimensions 300 • 130 • 60 cm (W • H • D)
Differential pressure measurement 0 – 3000 Pa
Aerosols Liquid aerosols (e. g. DEHS)
Aerosol concentrations Up to 106particles/cm3
Compressed air supply 10 barg
Pressure 1 – 7 barg relative