Search keywords:

product name, product type, model number,
test method, manufacturer, technique, application

Fabric Vertical Flame Spread Performance Tester

BACK
Share

Standards:

Applications:

Product Information:

The instrument is suitable for single-component or multi-component (coating, seam, multi-layer, mezzanine and similar combinations) of clothing, curtains and large tents, including arbor and door cover textile vertical flame spread performance determination. It is used to assess the performance of materials or combinations of materials under laboratory control conditions.

Standards:

GB/T 5456-2009 Textiles-Burning behavior-Flame spread properties of vertically oriented specimens

ISO 6941:2003 Measurement of flame spread properties of vertically oriented specimens

Features:

Electric wire ignition.

Burner automatic timing positioning

Ignition time 0 ~ 99.99 seconds arbitrarily set. Afterflame time, Afterglow time, flame spread time are automatically recorded, digital display.

Technical Parameters:

Items Parameters
Ignition time 0 ~ 99.99 seconds any set
Afterflame time, afterglow time, spread time 0 ~ 99.99 seconds, accuracy: ± 0.01 seconds
Flame height 10 ~ 60mm adjustable
Flame height measuring device  Scale height 40mm
Sample holder A height of 560mm rectangular frame
Gas Industrial propane or butane gas
Sample size 560mm × 170mm

FAQs:

1. What is a Fabric Vertical Flame Spread Performance Tester?

It is used to evaluate the flame spread and burning behavior of textile fabrics in a vertical position.

2. What fabrics can be tested?

It is suitable for testing clothing textiles, protective fabrics, upholstery, and other textile materials, depending on the applicable standard.

3. What does the test measure?

The tester can evaluate flame spread, afterflame time, afterglow time, and other combustion characteristics according to the relevant test method.

4. Which standards does it comply with?

The applicable standards depend on the specific test method and equipment configuration. Confirm the required standard before testing.

5. Why is vertical flame spread testing important?

It helps assess the fire performance of fabrics and supports textile safety evaluation, product development, and quality control.


Leave Message

Personal information

*
*
*
*

Information and comments