Medical Face Mask Fire Resistance Tester
BACKStandards:
GB 19083-2010 Technical requirements of medical protective masks YY 0469-2011 Technical requirements of medical surgical masks-Blood penetration test device GB 2626-2006 Respiratory protective products, Self - priming filtering particulates respirator-flammability BS EN136-1998 Respiratory protective de- vices -Full face masks - Requirements testing, marking
Applications:
The device is mainly used in the field of medical protection, and also can also be used for testing flame retardant properties of other textile coating materials.
Product Information:
This device is mainly used to test the flame retardant performance of medical masks. The combustion performance of the mask is measured through a head mold wearing a mask de-vice to contact with the flame at a certain linear speed.
Items | Parameters |
Linear velocity of head automatic movement | (60±5)mm/s |
Flame height | 40±4mm |
Flame temperature measurement | Thermocouple detection |
Burner height adjustment range | 30mm |
Metal head mold surface thickness | 2~3mm,Resistance to flame combustion |
Combustion timer | 0~9999.9s |
Accuracy | ±0.1s |
Power supply | AC220V,50Hz,500W |
Weight | About 30kg |
When the flame is (20±2)mm above the burner, the temperature is (800±50)℃. | |
distance between the top of the burner and mask lowest point is(20±2)mm. |