Mask synthetic blood penetration tester

BACK
Share

Standards:

GB 19083-2010 Technical requirements of medical face masks - Anti-penetration property of 5.5 synthetic blood. YY/T 0691-2008 Clothing for protection against infectious agents-Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected) YY 0469-2011 Technical requirements of medical face masks– Blood penetration test apparatus. ISO 22609:2004 Clothing for protection against infectious agents - Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected) ASTM F1862-07Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Know Velocity)

Applications:

The device is used for anti-synthetic blood penetration of medical surgical masks, and mainly used in the field of medical protection products. It can also be used for other textile coating material determination.

Product Information:

Medical Face Mask Synthetic Blood Penetration Tester is used for evaluating the resistance to synthetic blood penetration (splash resistance) of medical face masks. The test method challenges medical face masks with a fixed volume of synthetic blood directed at high velocity at the center of the mask.

 
Application:
The device is used for anti-synthetic blood penetration of medical surgical masks, and mainly used in the field of medical protection products. It can also be used for other textile coating material determination.
 
Features:
1.  The convex sample fixing device can simulate the actual state of use of a mask, leave a test target area without damaging the sample, and distribute the synthetic blood in the target area of the sample.
2.  The special constant pressure spray device can spray a certain volume of synthetic blood in a controlled time. 3. It can fully simulate the human body's average blood pressure of 10.6kPa, 16kPa, and 21.3kPa spray speeds for testing.
3.  Equipped with a target plate, which can block the high-pressure edge portion of the sprayed liquid flow, and only let the steady-state flow portion be sprayed on the sample, which increases the accuracy and repeatability of the liquid velocity sprayed on the sample.
4. Color touch screen operation.
5. Equip with thermal printer for printing results.
 
Parameter:
Spray Distance:300mm+/-10mm
Spray Speed:450cm/s,550 cm/s,635 cm/s
Nozzle:Diameter: 0.8mm, Length::12.7mm
Spray time:0.01s-99.99s
Synthetic blood volume:2L
Power Supply:AC 220V  50Hz
Mask Synthetic Blood Penetration Tester