Footwear Slip Resistance Tester
BACKStandards:
Applications:
This equipment is designed to perform dynamic tensile testing (or De Mattia flex cracking testing) on rubber specimens under both high- and low-temperature conditions.
For dynamic tensile testing, the device is equipped with a specimen break detection function, using a force sensor as the detection method. Compared with traditional detection methods, it offers advantages such as enhanced safety and reliability, no noise, and no requirement for specific lighting conditions.
Product Information:
Technical Parameters
Control Mode: Dedicated computer software + PLC
Number of Test Stations: 24 stations / 12 stations
Test Cycles: 0–9.999.999. freely set within the range
Test Functions: Dynamic tensile fatigue test and De Mattia flex cracking test
Dynamic Tensile Fatigue Test:
Clamp distance manually adjustable according to specimen length
Reciprocating stroke set in the program and automatically adjusted
24 stations with independent counting; counting stops when a specimen breaks; the machine stops automatically when all specimens break, or when the preset number of cycles is reached (selectable stop mode)
Break detection method: force sensor
De Mattia Flex Cracking Test:
Equipped with a 75 mm standard fixture; clamp distance manually adjustable
Default reciprocating stroke of 57 mm, automatically controlled
Unified counting for 24 stations
Observation button available; automatic positioning to observation distance: 65.0 ± 0.5 mm
Test Modes: Fixed-value test and programmed test
Fixed-value test: Constant temperature and constant speed
Programmed test: Segmented settings for temperature, speed, number of cycles, holding temperature, and holding time (up to 10 segments)
Test Speed: 20–300 cycles/min, freely adjustable
Test Temperature: -50°C to 180°C, freely adjustable
Cooling Method: Two-stage system with fully enclosed imported compressor refrigeration
Temperature Control: Anti-integral saturation PID with BTC balanced temperature control system; built-in circulation fan for lateral air circulation
Temperature Fluctuation: ±0.5°C
Temperature Deviation:
0°C to 100°C: ≤1°C
-50°C to 0°C and 100°C to 180°C: ≤2°C
Note: Measurement taken at the 24 specimen positions (entire test fixture)
Temperature Uniformity:
0°C to 100°C: ≤±1°C
-50°C to 0°C and 100°C to 180°C: ≤±2°C
Note: Measurement taken at the 24 specimen positions (entire test fixture)
Distance Between Upper and Lower Clamps: 0–80 mm, manually adjustable
Reciprocating Stroke of Moving Clamp: 10–80 mm, freely adjustable with automatic control
Overall Dimensions (W × D × H, approx.): 3500 mm × 2700 mm × 2300 mm
Power Supply: AC 380V, 10 kW, 50 Hz
Main Accessories
Main unit – 1 set
Dedicated testing software – 1 set
Product instruction manual – 1 copy