Search keywords:
product name, product type, model number,
test method, manufacturer, technique, application
G-ZC5361H Tracheal tube collapse resistance tester
BACKStandards:
ISO 5361:2023 Annex H
Applications:
Product Information:
1. Transparent tube: Φ8 mm polyurethane tube, its length is approximately twice the effective lengthof the cuff, and its inner diameter is twice the nominal outer diameter of the test endotracheal tube,with a deviation within 5%.
2. Water bath: Temperature controlled at (40±1)°C.
3. Gas supply source: ≤ 0.7 MPa (gas source provided by user)
4. Pressure indicator: 0–10 kPa; 0–1.45 psi (standard configuration), with an accuracy of ±5%.
5. Steel ball: Diameter is 75% of the nominal inner diameter of the test endotracheal tube.
Product principle
After the cuff of the endotracheal tube is inflated in a transparent tube, a steel ball is passed through its lumen to measure the collapse caused by the pressure inside the cuff.
Technical Parameters
• 7-inch high-definition color LCD screen, displaying test data and curves in real time.
• Pressure and time can both be preset; simply input the values to enter the test mode.
• The device has a powerful built-in data storage function, capable of storing up to 100.000 data entries.
• Automatic result determination; test data can be printed.
• Supports quick viewing and printing of historical data.
• Based on a flat design concept, the UI has been redesigned, providing users with a simpler, more efficient, and comfortable operating experience.
• Energy-saving and environmentally friendly design; aesthetically pleasing; ultra-low power consumption meets national energy-saving and environmental protection requirements.
• Standardized, modular, and serialized design concepts maximize the satisfaction of users' personalized needs.