Search keywords:
product name, product type, model number,
test method, manufacturer, technique, application
Liquid Sliding Penetration Comprehensive Performance Tester
BACKStandards:
Applications:
The tester is widely used in:
(1) Diaper and hygiene product manufacturing for evaluating absorbency, strike-through, and backflow resistance
(2) Non-woven fabric R&D for assessing liquid penetration properties of sanitary materials
(3) Quality control laboratories for verifying batch-to-batch performance of diapers and pads
(4) Third-party testing institutions for product performance certification
(5) Regulatory and supervisory agencies for monitoring compliance with hygiene and safety standards
Product Information:
The G373 Liquid Sliding Penetration Comprehensive Performance Tester is a precision instrument designed to evaluate the liquid penetration performance of diapers and hygienic non-woven fabrics. It accurately measures strike-through, back-seepage, and overall penetration efficiency, simulating real-world liquid exposure to assess product quality and reproducibility. The tester integrates a solenoid-controlled funnel, adjustable sample holder, and automated data acquisition for precise and repeatable results.
Standards
QB/T 2493-2000: Diapers (including diaper care/pads)
FZ/T 60017-1993: Hygienic thin non-woven fabric liquid permeability test method
ISO 9073-6: Nonwovens – Determination of liquid penetration
ASTM F1869/F1869M: Standard test method for liquid penetration resistance of textiles
Product Features
(1) Solenoid-controlled funnel: Simulates liquid application for penetration and strike-through testing
(2) Adjustable mechanical sample holder: Supports multi-test setups with precise angle control for reproducibility
(3) Sensor-integrated control system: Captures real-time liquid penetration data
(4) Data processing software: Transfers test results to a computer, generates curves, and enables in-depth analysis
(5) Automated operation: Simplifies the test process for consistent and accurate results
Technical Specifications
| Parameter | Specification |
|---|---|
| Sample Holder Angle | 30° ± 2° |
| Penetrating Efficiency | Time to penetrate 50 mL of liquid |
| Pressing Weight | Ø100 mm, 1.2 ± 0.002 kg (produces 1.5 kPa pressure) |
| Pressing Block Pressurization Time | 1 min |
| Funnel Flow | 80 mL / 6 s |
| Strike-Through Penetration Time | 10 min |
| Power Supply | 220V, 50Hz |
| Instrument Weight | 30 kg |
| Dimensions | 500 × 600 × 500 mm (L×W×H) |
Accessories
Standard sample holder with adjustable angle
Solenoid-controlled funnel
Pressing block (Ø100 mm)
Data acquisition and software package
Calibration tools
Test Procedures
(1) Place the diaper or non-woven sample on the sample holder and adjust to the specified angle
(2) Fill the funnel with the test liquid and set the solenoid valve for controlled flow
(3) Start the test: the liquid penetrates the sample, and sensors capture real-time data
(4) Apply the pressing block for 1 minute if required to simulate pressure
(5) Record strike-through, back-seepage, and penetration data; analyze via software
(6) After testing, clean the funnel, sample holder, and pressing block to avoid residue buildup
(7) Periodically calibrate the flow rate, angle, and sensor system to ensure consistent and accurate results