|

Search keywords:

product name, product type, model number,
test method, manufacturer, technique, application

Thermal Conductivity Tester (Guarded Hot Plate Method)

BACK
Share

Standards:

Applications:

Product Information:

Thermal Conductivity Tester adopts the internationally recognized hot plate method for measuring thermal conductivity and thermal resistance. It features computer-controlled fully automatic testing, generating comprehensive experimental reports. The instrument is suitable for testing the thermal performance of low-thermal-conductivity materials such as insulation materials and refractory materials. With an additional sample frame, it can also test powder materials. It is widely used in universities, research institutes, quality inspection centers, and industrial enterprises.

Reference Standards

The instrument is designed in accordance with the following standards:

GB/T 10295-2008 and ISO 8301-1991 – Determination of steady-state thermal resistance and related properties of insulating materials by the guarded hot plate method.

ASTM C518-04 – Test method for steady-state thermal flux and thermal transmission characteristics using a heat flow meter.

EN 12667-2001 and EN 12939-2001 – Thermal performance of building materials and products; determination of thermal resistance using the guarded hot plate and heat flow meter method.

System Composition

Hot Plate Heater

High-precision programmable temperature control.

Temperature measurement resolution on the hot surface: 0.01°C.

Cold Plate

Temperature-controlled constant-temperature water bath.

Temperature fluctuation: <0.05°C.

Temperature measurement resolution: 0.01°C.

Protective Control Device

Effectively eliminates environmental influences on the test.

Ensures measurement accuracy and repeatability.

Optional Imported Heat Flow Meter

High stability and reliable performance.

Main Specifications

Thermal Conductivity Testing Range: 0.0010–1.0 W/m·K,Resolution: 0.0005 W/m·K

Hot Plate Temperature Range: Room temperature – 100°C,Resolution: 0.01°C

Cold Plate Temperature Range: 0–99.99°C,Resolution: 0.01°C

Sample Size: 600 × 600 × (10–100) mm

Testing Accuracy: ±3%

Repeatability: ±1%

Automatic Thickness and Pressure Functions,Supports automatic thickness measurement and reading.

Automatic pressure application and reading.Electric loading and unloading with adjustable speed.

Measured material pressure can be set.

Transparent Acrylic Protective Cover.Side-opening door allows observation of the testing process.

Optional Sample Frame,Enables measurement of powder materials.

The pressure application is smooth, allowing measurement of thermal conductivity under different pressure conditions.Equipped with a geared motor for automatic loading.High-precision pressure sensor automatically measures applied pressure.Pressure measurement range: 0–1000 N.

Thickness Measurement.High-precision displacement sensor.Measurement range: 0–100.00 mm.

Hot and Cold Plate Heat Transfer Area: 600 × 600 mm

Power Supply: 220 V, 50 Hz

Instrument Configuration

DRPL-III-600 Thermal Conductivity Tester Main Unit: 1 set

Testing Software (with communication interface): 1 set

Constant-Temperature Water Bath and Controller: 1 set

User-Provided Computer

Leave Message

Personal information

*
*
*
*

Information and comments