Search keywords:
product name, product type, model number,
test method, manufacturer, technique, application
G-0461E Bending airflow resistance increase tester
BACKStandards:
ISO 5367-2023. YY/T 0461-2003. YY/T 0753.1-2009. YY/T 1543.
Applications:
Product Information:
This product is suitable for testing the performance of breathing tubing used in anesthesia machines and ventilators,primarily referencing standard ISO 5367-2023. YY0461 Appendix E, Test for Increased Airflow Resistance in Bending. It is suitable for use by quality inspection agencies, medical device testing institutes, research institutions,medical device manufacturers, and other organizations.
Test Principle
Bending airflow resistance: Based on the completed airflow resistance test, suspend the tubing for the bending test on a small-diameter metal column. Tie a weight to the end of the tubing to ensure the breathing tubing is in contact with half the circumference of the metal column. Introduce air into the tubing at the manufacturer's rated flow rate and record the pressure increment.
Technical parameters
| Item | Specification |
|---|---|
| Pressure Range | 0~600 KPa; 0~87 psi (standard configuration) |
| Flow Rate Range | 0~20 L/min |
| Inflation Head Specifications | Inner diameter Φ5.5~8.5 mm (standard configuration); other specifications can be customized |
| Air Source Pressure | 0.25 MPa~0.7 MPa (air source to be provided by user) |
| Air Source Interface | Φ8 mm polyurethane tubing |
| Electrical Control Box Dimensions | 325 mm (L) × 430 mm (W) × 205 mm (H) |
| Power Supply | AC 220V, 50Hz |
| Net Weight | Electrical control box: 10 kg |