Search keywords:

product name, product type, model number,
test method, manufacturer, technique, application

Guide Wire Flexibility Tester

BACK
Share

Standards:

Applications:

Product Information:

This instrument is suitable for flexibility and stiffness testing of the distal tip and shaft of non-vascular and intravascular catheter accessories, fallopian tube catheters, high-pressure angiographic catheters, microcatheters, balloon dilation catheters, guide wires, etc.It is designed and manufactured in accordance with relevant clauses of three-point bending test methods specified in YY/T 1554. YY 0450.1. and YY/T 0858-2011. and is widely used by medical device manufacturers and testing institutions.

Standards

YY/T 1554. YY 0450.1. YY/T 0858-2011. etc.

Features

Four-level user permission management ensures data integrity and standardization.

Customizable permission control allows flexible assignment of user roles and strict operation restrictions.

Comprehensive password protection prevents unauthorized operations and ensures data traceability.

Encrypted data storage ensures completeness, security, and reliability of all test records.

Large LCD display shows the testing process; PVC operation panel is user-friendly and practical.

Equipped with a globally recognized force measurement system for better repeatability and flexibility.

High-precision displacement sensor ensures accurate speed transmission.

Industry-leading multi-function capability: tip flexibility, three-point bending, and catheter stiffness testing in one machine.

Built-in standard test programs with one-button operation and automatic test completion judgment.

Integrated micro-printer outputs complete test information including device ID, sample batch number, operator, results, and test time.

System supports ISP online upgrade for customized services.

Specifications

Parameter

Specification

Measurement range

0–10 N (customizable)

Resolution

±0.001 N

Test speed

0–10.0 mm/min (stepless speed regulation)

Guide hole diameter

4 mm (customizable)

Feeding stroke

1–100 mm (adjustable)

Users

50 (usernames can include numbers, letters, symbols)

Data interface

RS232 (connectable to LIMS system)

Dimensions (L×W×H)

500 × 480 × 475 mm

Weight

28 kg

Environmental Requirements

Parameter

Specification

Working temperature

5°C–40°C

Relative humidity

25%–80% (non-condensing)

Power supply

220V, 50Hz

Configuration

Standard configuration: Main unit, Micro printer, Touchscreen LCD, Test fixtures.

Leave Message

Personal information

*
*
*
*