Temperature and Humidity Test Chamber
BACKStandards:
Applications:
Product Information:
Parameters
Temperature range : -50℃~+120℃
Temperature deviation : ≤±2.0℃(the full load)
Temperature fluctuation: ≤±0.5℃(the full load)
Temperature uniformity: ≤2℃(the full load)
Sensor : Temperature sensors: 3. one for box overheating, one for box temperature monitoring, and one for component temperature monitoring
Humidity sensor : 1. wet dry bulb comparison method
Temperature control :Air temperature control at the air outlet
Heating and cooling rate:1.5~3.3 ℃/min @ full load, linear
Humidity range : 20%~95%R·H
Humidity deviation : ≤±2%RH(the full load)
Humidity fluctuation: ≤±1.5%R·H(the full load)
Humidity uniformity: ≤±3%R·H(the full load)
Test program : 25℃、65℃/85%R.H、85℃/85%R.H、90℃/90%R.H、95℃/95%R.H、105℃
Internal dimensions of the chamber: 2.0 m (L) × 1.5 m (W) × 2.0 m (H).
Chamber structure
Observing from the front door, the workspace is located in front, while the refrigeration unit and electrical control cabinet are located behind. The electrical control cabinet and refrigeration room are independently enclosed from each other. The air duct in the studio is equipped with an evaporative dehumidifier, heater, humidification inlet pipe, and fan. The air circulation is designed
with up and out air and down return air. The bottom plate of the studio is equipped with a drainage ditch, which can smoothly discharge the condensed water from the floor inside the box to the outside.