NewsInformation Center
Home>News > > How to Use a Medical Injection Needle Tubing Rigidity Tester?

How to Use a Medical Injection Needle Tubing Rigidity Tester?

2026/11/05

Share: 

Share
The Medical Injection Needle Tubing Rigidity Tester is a specialized instrument designed to evaluate the bending resistance and deformation capability of medical injection needle tubing. It quantitatively determines whether the needle tubing is sufficiently rigid and whether it is likely to bend during puncture. This test is essential for ensuring the mechanical performance and reliability of medical injection needles.So, how is a Medical Injection Needle Tubing Rigidity Tester used? As a device designed to measure the bending deformation of needle tubing, it is common for users to be unfamiliar with its operation after purchasing the equipment. The following sections introduce its core functions, working principle, and specific operating procedures to help users better understand how to operate the instrument.

Core Functions of the Medical Injection Needle Tubing Rigidity Tester

1. Efficient Detection of Mechanical Rigidity

The Medical Injection Needle Tubing Rigidity Tester is primarily used to evaluate the mechanical rigidity of medical injection needle tubing, ensuring that the needle tubing is less likely to bend or break during use, thereby safeguarding patient safety and ensuring effective performance.

By applying a specified force to the needle tubing, the instrument accurately measures its deformation under load to determine whether its rigidity complies with the applicable standard requirements.

2. Intelligent Operation and Data Processing

The tester adopts a Linux operating system and is equipped with a 7-inch bilingual touchscreen supporting Chinese and English, providing smooth and convenient operation.

It supports bilingual input, allowing users to freely edit and customize information such as the manufacturer, product name, product batch, test personnel, and various product performance parameters.

The instrument also incorporates PLC and ARM-based high-efficiency data processing technology, enabling real-time display of loading data, improved testing efficiency, and more intuitive results.

3. Self-Protection and Warning Functions

The Medical Injection Needle Tubing Rigidity Tester features comprehensive self-protection mechanisms, including sensor protection and automatic power-off protection against overload.

In the event of improper operation or instrument overload, the sensor protection mechanism is triggered to effectively reduce maintenance costs and protect both the instrument and the operator.

Additionally, when a test result fails to meet the specified requirements, the instrument activates an audible buzzer alarm to alert the testing personnel and prompt timely corrective action.

Working Principle of the Medical Injection Needle Tubing Rigidity Tester

The tester operates primarily based on constant-force testing technology.

During testing, the medical injection needle tubing specimen is first properly positioned and secured. The instrument then applies a predetermined force to the needle tubing through its built-in loading system at a uniform and stable rate, ensuring consistent force application without fluctuations and preventing specimen damage or test data distortion caused by sudden changes in loading.

Throughout the loading process, the instrument continuously detects the deformation of the medical injection needle tubing. Its built-in high-precision measurement module accurately converts the deformation data into readable and analyzable electrical signals.

The control system then calculates and processes the acquired data to determine the rigidity value of the medical injection needle tubing, enabling precise evaluation of the specimen's rigidity performance and ensuring that the test results comply with applicable industry standards and product quality requirements.

Operating Procedure of the Medical Injection Needle Tubing Rigidity Tester

1. Preliminary Test Preparation

Prepare all required testing accessories, including fixtures, a displacement micrometer, weights, and the specimens to be tested.

Carefully inspect all accessories to ensure that they are free from deformation or damage. Confirm that the specimen specifications meet the testing requirements and ensure that all items are clean and free from contaminants to provide suitable conditions for the subsequent test.

2. Test Parameter Setting

After starting the instrument, enter the operation interface on the touchscreen.

According to the specifications and testing requirements of the medical injection needle tubing, accurately set the key test parameters, including the specified load and maximum deflection.

After completing the parameter settings, carefully verify all values and save them once confirmed to be correct, preventing deviations in test results caused by incorrect parameter settings.

3. Micrometer Installation and Adjustment

Place the dedicated fixture securely in the designated position on the instrument.

Then install the displacement micrometer accurately into the corresponding interface of the fixture. Adjust its position to ensure secure installation without looseness.

Calibrate the micrometer to its zero position to ensure the accuracy of subsequent test measurements.

4. Specimen Clamping and Fixation

Place the specimen to be tested securely into the groove of the fixture.

Adjust the specimen position to ensure that it fits properly against the fixture without displacement or distortion.

Then operate the fixture to clamp and secure the specimen firmly, preventing movement during testing and ensuring measurement accuracy.

5. Initiating the Test Procedure

After confirming that specimen clamping, parameter settings, and micrometer adjustment are all correct, click the "Start Test" button on the touchscreen.

The instrument will initiate the test program, and the displacement micrometer will move downward slowly at the preset speed, maintaining uniform and stable movement throughout the process.

Avoid excessive movement speed, which may damage the specimen or cause test data distortion.

6. Loading Control

As the micrometer moves downward, it gradually applies a uniform force to the specimen.

During loading, continuously monitor the load value displayed on the touchscreen and carefully observe the specimen condition to prevent abnormal situations, such as specimen fracture or fixture loosening.

7. Test Completion and Data Recording

Continue monitoring the load value until it stabilizes and no longer fluctuates significantly.

The instrument will automatically record the test data and generate the test results.

After verifying that the recorded values are correct, save the data and activate the printing function to print the test report.

Once the report has been printed, click the "End Test" button to terminate the test program and complete the testing procedure.

In summary, operating the Medical Injection Needle Tubing Rigidity Tester requires strict compliance with the specified test parameters and appropriate specimen preparation. The testing process must be accurate and reliable, and the results should be carefully analyzed. Test data should also be promptly maintained, organized, and summarized to improve work efficiency and ensure data reliability.

Conclusion

With continuous technological advancement, needle tubing rigidity testers are constantly evolving and improving.Modern testers significantly improve measurement accuracy through the adoption of more advanced sensors and algorithms. Meanwhile, next-generation instruments generally offer multiple testing modes to accommodate different types and specifications of needle tubing.Furthermore, with the integration of artificial intelligence and Internet of Things (IoT) technologies, modern needle tubing rigidity testers can support automated testing, remote monitoring, and cloud-based data storage, substantially improving testing efficiency and operational convenience.

Previous: What are the advantages of a suture needle cutting force tester?
 N e x t   : the last page