Electrostatic Charge Decay Meter is a kind of static electricity that tests the electrostatic diffusivity of various materials Attenuation testing instrument, the sample is irradiated by air ions caused by halo discharge Shooting and charging, after the air ion irradiation stops, the charge is measured by the measuring device Attenuation curve, after which the half-life of voltage attenuation is automatically calculated by the analyzer.

Use function of Shishido Static Honestmeter:

1. Setting of attenuation rate (half-life is usually set at 50%): 2 ~ 100%
2. Computer data storage of test data
3. The voltage coordinates and time coordinates can be enlarged or reduced at will
4. The temperature and humidity data is automatically saved to the folder

Shishido Static Honestmeter

We all know that static electricity is not allowed in some jobs, and the static electricity on workers must be eliminated before working. However, whether static electricity is completely eliminated or not can be determined by static attenuation tester. However, there are still some points to pay attention to when using the electrostatic attenuation tester.

1. The tester is specially used for static electricity treatment, and needs to be away from water, oil, solvent and other conductive pollutants.

2. Do not insert any object into the opening for fixing the sensor, and do not let any foreign matter enter the opening of the sensor.

3. Please ground the instrument correctly for the measurement of the instrument.

4. This instrument is an electronic device, and its sensors may be very sensitive to mechanical vibration and impact; Of course, it also contains a micro processing chip and electronic circuit, so it can't be used in the environment with many electronic noises.

5. The Static Honest Meter should be stored and used in an environment with humidity less than 60%, and special care should be taken to ensure that no dew is formed in the environment, otherwise it will cause electrical insulation failure of the product.

6. Do not operate in corrosive acid, alkali, smoke or corrosive gas such as chlorine.

7. If the instrument falls, it is easy to be damaged. If this happens, the electrostatic attenuation tester should be carefully checked and repaired by authorized technicians.