Home>News > > What are the correct steps for using an ozone aging tester?

What are the correct steps for using an ozone aging tester?

2023/02/13

Share: 

Share
The correct ozone aging test chamber use steps can extend the service life of the equipment, tWhat are the correct steps for using an ozone aging tester?
 
A, ozone aging test chamber timing settings: set through the timer panel "+" "-" button to set (H: hours, M: minutes, S: seconds), such as 2 hours and 30 minutes, set "02H30", set up to open the timing switch, press the power "ON" button, this time the equipment into high temperature constant temperature operation, the time is up, the equipment will automatically cut off the power.
Ozone aging chamber
Second, ozone aging test chamber ozone test start: open the green button power. Adjust the temperature of the test chamber to 40 degrees, and so the test chamber temperature stability. The ozone meter to 0, and so 20 minutes after the ozone button to open. Instrument settings see the instrument manual; humidity test, first check whether there is water inside the tank, no water to add water. Turn on the hot and humid button. The test instrument is adjusted to the required test temperature, open the refrigeration button.
 
Third, ozone aging test chamber dynamic test: open the turntable button and stretch button, according to the need to adjust the speed, the exhaust button is waiting for the end of the test when the ozone discharge to the outdoor open.
 
Fourth, ozone aging test chamber alarm temperature setting: press the instrument setting key "SET" for 5 seconds to enter the internal menu of the instrument:
 
1.AL1 (alarm 1): generally AL1 = SV + 15 ℃, the role of the alarm when the high temperature, low temperature does not need to be set. Example: SV is 50℃, then AL1 is 65℃.
 
2. ATU (self-tuning): self-tuning as 0000 means closed, 0001 is open. This machine is generally opened without self-tuning, if a set temperature value is difficult to be constant, then open self-tuning. In the process of self-tuning, the temperature inside the box will fluctuate, is a normal phenomenon.
 
The above on the ozone aging test chamber timing, start, dynamic experiments and alarm temperature four steps for a brief introduction, I hope to help customers who buy ozone aging test chamber.